Pipete Research: date tehnice |
Research monocanal cu volum fix: date tehnice| Volum | Inexactitate* | Imprecizie* | | Buton de operare Gri, pentru varfuri de 20 µl sau GELoader | | 1 µl | ±2.5% | ≤1.8% | | 2 µl | ±2.0% | ≤1.2% | | 5 µl | ±1.5% | ≤0.8% | | 10 µl | ±1.0% | ≤0.5% | | Buton de operare Galben, pentru varfuri de 200 µl | | 10 µl | ±1.0% | ≤0.5% | | 20 µl | ±0.8% | ≤0.3% | | 25 µl | ±0.8% | ≤0.3% | | 50 µl | ±0.7% | ≤0.3% | | 100 µl | ±0.6% | ≤0.2% | | Buton de operare Albastru, pentru varfuri de1,000 µl | | 200 µl | ±0.6% | ≤0.2% | | 250 µl | ±0.6% | ≤0.2% | | 500 µl | ±0.6% | ≤0.2% | | 1.000 µl | ±0.6% | ≤0.2% | | Buton de operare Rosu, pentru varfuri de 2,500 µl | | 1.500 µl | ±0.6% | ≤0.2% | | 2,000 µl | ±0.6% | ≤0.2% | | 2,500 µl | ±0.6% | ≤0.2% |
| | Lichiid: Apa bidistilata, degazata | | Temperatura: 20 - 25 °C, constanta in limita a ±0,5 °C | | Nr. de masuratori: 10, in conformitate cu EN ISO 8655 folosind varfuri originale Eppendorf® |
* Datele privind imprecizia (eroare de masurare aleatorie) si inexactitatea (eroare de masurare sistematica) conform EN ISO 8655 sunt valabile numai cand se folosesc varfuri originale Eppendorf. Research monocanal cu volum variabil: date tehnice| Interval de volum | Volum | Inexactitate* | Imprecizie* | | Buton de operare Intunecat, pentru varfuri de 10 µl | | 0.1 - 2.5 µl | 0.25 µl | ±12.0% | ≤6.0% | | 1.25 µl | ±2.5% | ≤1.5% | | 2.5 µl | ±1.4% | ≤0.7% | | Buton de operare Gri, pentru varfuri de 20 µl | | 0.5 - 10 µl | 1 µl | ±2.5% | ≤1.8% | | 5 µl | ±1.5% | ≤0.8% | | 10 µl | ±1.0% | ≤0.4% | | Buton de operare Galben, pentru varfuri de 200 µl | | 2.0 - 20 µl | 2 µl | ±5.0% | ≤1.5% | | 10 µl | ±1.2% | ≤0.6% | | 20 µl | ±1.0% | ≤0.3% | | | | 10 - 100 µl | 10 µl | ±3.0% | ≤1.0% | | 50 µl | ±1.0% | ≤0.3% | | 100 µl | ±0.8% | ≤0.2% | | | | 20 - 200 µl | 20 µl | ±2.5% | ≤0.7% | | 100 µl | ±1.0% | ≤0.3% | | 200 µl | ±0.6% | ≤0.2% | | Buton de operare Albastru, pentru varfuri de 1,000 µl | | 100 - 1,000 µl | 100 µl | ±3.0% | ≤0.6% | | 500 µl | ±1.0% | ≤0.2% | | 1,000 µl | ±0.6% | ≤0.2% | | Buton de operare Mov, pentru varfuri de 5 ml | | 500 - 5,000 µl | 500 µl | ±2.4% | ≤0.6% | | 2,500 µl | ±1.2% | ≤0.25% | | 5,000 µl | ±0.6% | ≤0.15% |
* Datele privind imprecizia si inexactitatea sunt valabile numai cand se folosesc varfuri originale Eppendorf. Research multicanal cu volum variabil: date tehnice Interval de volum (µl) | Varf Eppendorf | Volum (µl) | Inexactitate* (%) | Imprecizie* (%) | 0.5-10 | Cristal 20 | 1 | ±8.0 | ≤5.0 | | | 5 | ±4.0 | ≤2.0 | | | 10 | ±2.0 | ≤1.0 | 10-100 | 100 | 10 | ±3.0 | ≤2.0 | | | 50 | ±1.0 | ≤0.8 | | | 100 | ±0.8 | ≤0.3 | 30-300 | 300 | 30 | ±3.0 | ≤1.0 | | | 150 | ±1.0 | ≤0.5 | | | 300 | ±0.6 | ≤0.3 |
* Datele privind imprecizia si inexactitatea sunt valabile numai cand se folosesc varfuri originale Eppendorf.
|